Safety and performance of a cell-impermeable endoprosthesis for hemodialysis vascular access outflow stenosis: a brazilian multicenter retrospective study

dc.contributor.authorHarduin, Leonardo de Oliveira
dc.contributor.authorBarroso, Thiago Almeida
dc.contributor.authorGuerra, Julia Bandeira
dc.contributor.authorFilippo, Márcio Gomes
dc.contributor.authorAlmeida, Leonardo Cortizo de
dc.contributor.authorVieira, Brunno Ribeiro
dc.contributor.authorMello, Renata Silveira
dc.contributor.authorGalhardo, Adriano Martins
dc.contributor.authorMatos, Jorge Paulo Strogoff de
dc.date.accessioned2025-12-12T15:36:04Z
dc.date.issued2024
dc.description.abstractObjective: To evaluate the safety and performance of WrapsodyTM, a cell-impermeable endoprosthesis (CIE), for treating hemodialysis vascular access outflow stenosis. Methods: Investigators retrospectively analyzed 113 hemodialysis patients treated with a CIE (11/ 2021–12/2022) across four centers in Brazil. De novo or restenotic lesions were treated. The primary efficacy outcome measure was target lesion primary patency (TLPP) at 1, 3, 6, and 12 months; the primary safety outcome measure was the absence of serious local or systemic adverse events within the first 30 days post-procedure. Secondary outcome measures included technical and procedural success, access circuit primary patency (ACPP), and secondary patency at 1, 3, 6, and 12 months post-procedure. Results: Thirty-nine patients (34.5%) had thrombosed access at the initial presentation, and 38 patients (33.6%) presented with recurrent stenosis. TLPP rates at 1, 3, 6, and 12 months were 100%, 96.4%, 86.4%, and 69.7%, respectively. ACPP rates were 100% at 1 month, 89.2% at 3 months, 70.9% at 6 months, and 56.0% at 12 months. The target lesion secondary patency rates at 1, 3, 6, and 12 months were 100%, 97.3%, 93.6%, and 91.7%, respectively. In the adjusted multivariate Cox regression analysis, male sex and endoprosthesis with diameters of 10, 12, 14, and 16 mm were associated with improved primary patency rates. No localized or systemic serious adverse event was observed through the first 30 days postprocedure. Conclusion: The CIE evaluated in this study is safe and effective for treating peripheral and central outflow stenoses in hemodialysis vascular access. Level of Evidence Level 2b, cohort study.
dc.identifier.citationHARDUIN, Leonardo de Oliveira; BARROSO, Thiago Almeida; GUERRA, Julia Bandeira; FILIPPO, Márcio Gomes; ALMEIDA, Leonardo Cortizo de; VIEIRA, Brunno Ribeiro; MELLO, Renata Silveira; GALHARDO, Adriano Martins; MATOS, Jorge Paulo Strogoff de. Safety and performance of a cell-impermeable endoprosthesis for hemodialysis vascular access outflow stenosis: a brazilian multicenter retrospective study. Springer. [S. l.]. v.47. p1057–1065. 2024 DOI https://doi.org/10.1007/s00270-024-03790-1
dc.identifier.urihttps://ortobase.into.saude.gov.br/handle/123456789/288
dc.language.isoen
dc.publisherSpringer
dc.subjectHemodialysis
dc.subjectCell-Impermeable endoprosthesis
dc.subjectCovered Stent
dc.subjectStent Graft
dc.subjectVascular Access
dc.subjectStenosis
dc.titleSafety and performance of a cell-impermeable endoprosthesis for hemodialysis vascular access outflow stenosis: a brazilian multicenter retrospective study
dc.typeArticle

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Harduin, Leonardo de Oliveira et al_2024.pdf
Size:
857.98 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed to upon submission
Description: